Solids
Advanced Hall magnetometer with improved performance
Kh. R. Rostami Kotelnikov Institute of Radioengineering and Electronics, Fryazino Branch, Russian Academy of Sciences
Abstract:
A sharp jump in the magnetic response near the first critical twins magnetic field $H_{ic1}$ was generated under the combined action of a constant biased and a local oscillating damped magnetic fields on the surface of the YBa$_2$Cu$_3$O$_{7-x}$ epitaxial film. The possibility of increasing due to this jump the sensitivity of a standard Hall magnetometer from $\sim$ 2.5 $\cdot$ 10$^{-3}$ Gs/Hz$^{1/2}$ to $\sim$ 8 $\cdot$ 10$^{-7}$ Gs/Hz$^{1/2}$ in the frequency range $\sim$ 1 kHz has been demonstrated. The range of the measured magnetic fields of the magnetometer was extended to the region of weak fields $\sim$ (8 $\cdot$ 10$^{-7}$–2.5 $\cdot$ 10$^{-3}$) Oe, while maintaining the accuracy and linearity of the order of $\sim$ 0.01%. To increase the sensitivity, linearity and noise sustainability the beginning of the operating point of the magnetometer was shifted to the twins values $H_{ic1}$. The increased spatial resolution of the magnetometer, determined by the linear dimensions of the YBa$_2$Cu$_3$O$_{7-x}$ film twins, was provided by an increase in the total magnetic field directed perpendicular to the film surfaces and the Hall transducer and reached $\sim$ 300 nm.
Keywords:
the twin high temperature superconductor, the density jump of the trapped magnetic flux, biased magnetic field, oscillating damped local magnetic bias field, two-stage Hall magnetometer.
Received: 04.01.2022 Revised: 28.04.2022 Accepted: 26.09.2022
Citation:
Kh. R. Rostami, “Advanced Hall magnetometer with improved performance”, Zhurnal Tekhnicheskoi Fiziki, 92:12 (2022), 1803–1813
Linking options:
https://www.mathnet.ru/eng/jtf7489 https://www.mathnet.ru/eng/jtf/v92/i12/p1803
|
| Statistics & downloads: |
| Abstract page: | 52 | | Full-text PDF : | 29 |
|