Zhurnal Tekhnicheskoi Fiziki, 2013, Volume 83, Issue 8, Pages 127–134
(Mi jtf8519)
|
|
|
|
This article is cited in 2 scientific papers (total in 2 papers)
Experimental instruments and technique
Optoelectronic system of the aerosol photometer in the detector of molecular condensation nuclei
V. D. Kuptsova, V. Ya. Katelevskiib, V. P. Valyukhova, E. N. Rybinb a Peter the Great St. Petersburg Polytechnic University
b OAO ENPO Neorganika, Elektrostal, Moscow oblast
Abstract:
We study the light scattering by aerosol particles and air in the photometer of the molecular condensation nuclei, as well as the sensitivity of the photodector of the photometer. The interference nature of light scattering by aerosol particles is established and is found to be comparable (in order of magnitude) with the scattering of light by air in the photometer. The sensitivity of the photometer can be increased by more than an order of magnitude due to optimization of the optoelectronic part of the photometer. The detection threshold for the target component of the gas analyzer is attained at the spontaneous ionization background level and not at the limiting sensitivity level of the photodetector.
Received: 18.10.2012
Citation:
V. D. Kuptsov, V. Ya. Katelevskii, V. P. Valyukhov, E. N. Rybin, “Optoelectronic system of the aerosol photometer in the detector of molecular condensation nuclei”, Zhurnal Tekhnicheskoi Fiziki, 83:8 (2013), 127–134; Tech. Phys., 58:8 (2013), 1211–1218
Linking options:
https://www.mathnet.ru/eng/jtf8519 https://www.mathnet.ru/eng/jtf/v83/i8/p127
|
|