Eurasian Journal of Mathematical and Computer Applications, 2016, том 4, выпуск 4, страницы 5–13
(Mi ejmca27)
|
|
|
|
Application of Reverse Time Migration (RTM) procedure in ultrasound tomography, numerical modeling
V. M. Filatova, В. В. Nosikova, L. N. Pestov Immanuel Kant Baltic Federal University, 236016 Kaliningrad, Russia
Аннотация:
This work is devoted to the medical ultrasound imaging problem. We present some results of numerical modeling, using Reverse Time Migration procedure. In the numerical experiment we use acoustical model that simulates cross-section of breast containing fat, glandular tissue and weak inclusions with a radius from 0.8mm. It is impossible to determine these inclusions on the image obtained after a standard procedure RTM with a constant speed of sound. We use an approach based on a modified version of RTM to determine unknown boundary separating fat and glandular tissue and weak inclusion. The final image contains all the inhomogeneities and close inclusions are separated at that.
Ключевые слова:
ultrasound medical tomography, Reverse Time Migration, numerical modeling.
Поступила в редакцию: 01.11.2016